Part numberType 8222Item numberBURK-8222HS code9032.89
Sensors & Instrumentation
Process Controllers
MOQ:1

Pricing

Request a quote

We reply within 24 hours

Product Description

Bürkert Type 8222 The Bürkert conductivity meter Type 8222 is designed for measuring the conductivity of liquids. The sensor element comprises a two-electrode cell and a Pt1000 temperature probe. The sensor itself is available with three different cell constants C. These are made of stainless steel for sensors with a constant C of 0.01 or 0.1 cm- 1 and of graphite for sensors with a constant C of 1.0 cm- 1. Thus, due to the measuring device’s design, Bürkert has simplified installation and maintenance work. The Type 8222 device is available in two variants. The first one, the so-called ELEMENT standard is proposed either with three adjustable outputs (two digital outputs and one analogue output) or with four adjustable outputs (two digital and two analogue) and can be equipped with a display. The display is only required for start-up, configuration (e.g. measuring range, units, calibration, limit values) or as a process value display. The second variant, the so-called ELEMENT neutrino, is a 2-wire device, without display, with a 4.20 mA current output or with a digital communication mode that can communicate either via IO-Link or in büS (Bürkert system bus based on CANopen). Both variants are available with a G 1½" union nut for installation in an adapter with a G 1½" external thread sensor connection that is connected to the process. The ELEMENT neutrino variant is also available with a G ¾" external thread for screwing into an adapter with a G ¾" internal thread sensor connection. The Type 8222 device converts the measurement signal and calculates the output signals, which are provided via one or two M12 plug connections in the ELEMENT standard variant or via an M12 plug connection in the ELEMENT neutrino variant or on a terminal strip via a cable gland. The device in the ELEMENT standard variant shows different values in different measuring units (if a display is installed).

Technical Specifications

HS code
9032.89
Medium temperature
-20–100 °C
Probe holder
PVDF, stainless steel 1.4571 (316Ti)
Electrode
For cell constant: • C = 0.01 cm- 1: stainless steel 1.4571 (316Ti) • C= 0.1 cm- 1: stainless steel 1.4571 (316Ti) • C = 1 cm- 1: graphite
Compatibility
With any pipe which is fitted with Bürkert S022 adapter. See data sheet Type S022  for more information.
Pipe diameter
DN 32...DN 110 (DN 06...DN 25 under specific conditions, i.e. when using reducers)
Probe
With cell constant • C = 0.01 cm- 1 • C= 0.1 cm- 1 • C = 1 cm- 1
Temperature sensor
Pt1000 integrated within the probe holder
Conductivity measurement
0.05 µS/cm...10 mS/cm
Temperature measurement
20...+ 100 °C (- 4...+ 212 °F)
Measurement deviation
± 3 % of measured value
Measuring range resolution
1 nS/cm
Power source (not supplied)
Limited power source according to UL/EN 62368‑1 standards or limited energy circuit according to UL/ EN 61010‑1 paragraph 9.4
DC reverse polarity protection
Yes
Overvoltage protection
Yes
Fluid temperature
Device with • G 1½ PVC union nut connection: 0...+ 50 °C (+ 32...+ 122 °F) • G 1½ PVDF union nut connection (on request for ELEMENT neutrino variant): - 20...+ 100 °C (- 4...+...
Process connection
G 1½ internal thread for use with Type S022 adapter See data sheet Type S022  for more information.
North America (USA/Canada)
UL Recognized for the USA and Canada
Ambient temperature
Operation and storage: - 10...+ 60 °C (+ 14...+ 140 °F)
Relative air humidity
≤ 85 %, without condensation
Height above sea level
Max. 2000 m
Operating condition
Continuous
Equipment mobility
Fixed
Application range
Indoor and outdoor Protect the device against electromagnetic interference, ultraviolet rays and, when installed outdoors, against the effects of climatic conditions.
Installation category
Category I according to UL/EN 61010‑1
Pollution degree
Degree 2 according to UL/EN 61010‑1
Cover
Polycarbonate (PC), transparent (opaque on request)
Housing
Stainless steel 1.4404 (316L), PPS

Standards & Compliance

HS Code
9032.89

Cross References

  • Burkert 8222
  • Buerkert 8222
  • Bürkert 8222
  • Type 8222
  • Burkert Type 8222

Product Documentation

Document preview: Bürkert Type 8222 Datasheet (EN)

Catalogue entry last verified: 2026-08-29. A manufacturer datasheet is attached to this entry — confirm application-critical values against it.

Frequently asked questions

Where can I buy Type 8222?
Type 8222 can be ordered from cballast.com as a Bürkert catalogue item: request a quotation on this page or email sales@cballast.com, and the order ships worldwide.
What is the manufacturer part number for this Bürkert product?
The manufacturer part number is Type 8222.
What is the HS code for Type 8222?
The customs tariff (HS) code for Type 8222 is 9032.89.
Who manufactures Type 8222?
Type 8222 is a Bürkert product in the Sensors & Instrumentation range.
Is a datasheet available for Type 8222?
Yes — the Type 8222 datasheet can be viewed and downloaded on this page.
What is the minimum order quantity for Type 8222?
The minimum order quantity is 1.
Is Type 8222 available?
Yes — Type 8222 is available to order: fulfilment is from supplier stock in cBallast's European network, dispatched directly from the warehouse to the customer, with the exact lead time stated in the quotation (reply within 24 hours).