The Type 8221 hygienic conductivity probes are used to determine electrical conductivity in a wide range of pure or concentrated liquids. Due to their hygienic and robust design, these conductivity probes are suitable for use in various application sectors, including the food & beverage, pharmaceutical, biotechnology and chemical industry. They are available in two technologies, with either 2 or 4 electrodes. Probes based on the 2-electrode principle are intended for use in pure liquids, particularly ultrapure water, as contamination affects the measurement. Probes based on the 4-electrode principle prevent polarisation phenomena and are not sensitive to contamination. The clever design guarantees an excellent linearity over the entire measurement range. All variants are fitted with an integrated temperature sensor (Pt1000) as standard. The probe has to be connected to the multiCELL transmitter/controller Type 8619.

Technical Specifications

Medium temperature
-20–150 °C
Seal
EPDM (conform to FDA ‑ 21CFR177.2600)
Surface quality
Ra < 0.4 μm (15 μin.), electro-polished (wetted metal surfaces)
Measuring element
4-electrode
Temperature sensor
Pt1000
Linearity 1.) (relative)
± 0.5...5 %
Degree of protection according
IP67, with connected device, inserted and screwed cable plug
Electrode
Stainless steel 316L/1.4404
Armature
PEEK (conform to FDA ‑ 21CFR 177.2415) and Stainless steel 316L/1.4404
Fluid temperature
20...+ 150 °C (- 4...+ 302 °F)
Process connection
1½" clamp connection • G 1 connection • G ¾ connection • NPT ¾ connection
Electrical connection
5-pin M12 male connector
Others
On request: 2-point calibration certificate
Ambient temperature
Operation: - 20...+ 150 °C • Storage: - 10...+ 60 °C
Probes
based on the 2-electrode principle are available in two electrode architectures: The hole is located 19 mm from the base of the armature. The electrode spacing is short and the...

Standards & Compliance

HS Code
9027.80

Cross References

  • Burkert 8221
  • Buerkert 8221
  • Bürkert 8221
  • Type 8221
  • Burkert Type 8221

Product Documentation

Document preview: Bürkert Type 8221 Datasheet (EN)

Catalogue entry last verified: 2026-08-13. A manufacturer datasheet is attached to this entry — confirm application-critical values against it.

Frequently asked questions

Where can I buy Type 8221?
Type 8221 can be ordered from cballast.com as a Bürkert catalogue item: request a quotation on this page or email sales@cballast.com, and the order ships worldwide.
What is the manufacturer part number for this Bürkert product?
The manufacturer part number is Type 8221.
What is the HS code for Type 8221?
The customs tariff (HS) code for Type 8221 is 9027.80.
Who manufactures Type 8221?
Type 8221 is a Bürkert product in the Sensors & Instrumentation range.
Is a datasheet available for Type 8221?
Yes — the Type 8221 datasheet can be viewed and downloaded on this page.
What is the minimum order quantity for Type 8221?
The minimum order quantity is 1.
Is Type 8221 available?
Yes — Type 8221 is available to order: fulfilment is from supplier stock in cBallast's European network, dispatched directly from the warehouse to the customer, with the exact lead time stated in the quotation (reply within 24 hours).