Hach HQ2200 Portable Multi-Meter kit pH — Sensor with 5 m Rugged Cable. Reference part number LEV015.98.22009.

Technical Specifications

Accessories Included
USB charger, power cable, and clips
Automatic Buffer Recognition
Yes
Backlight
Yes
Cable Length
1 m
Calibration Intervals/Alerts/Reminder
Off, selectable from 2 hours to 7 days
Certifications
CE, FCC, ISED, RCM, KC
ETL Verified
US DOE/ NRCan Energy Efficiency, RoHS
Data Export
USB connection to PC or USB storage device (limited to the storage device capacity).
Data Memory
10,000 data points
Data Storage
Automatic in Press to Read Mode and Interval Mode. Manual in Continuous Read Mode.
Dimensions (H x W x D)
63 mm x 97 mm x 220 mm
Display
Up to 3 parameters at a time, dependent on HQ model
Display Type
536x336 Mono-TFT
DO Measurement Range
0.1 - 20.0 mg/L (ppm), 1 - 200% saturation
DO Resolution
0.1
Enclosure Rating
IP67 (with battery compartment installed)
GLP Features
Date; Time; Sample ID; Operator ID, Calibration
Inputs
2
Instrument
Portable
ISE Direct Measurement Range
No
ISE Electrode Calibration
No
Kit?
Yes
Languages
English, Chinese - Simplified, Spanish, French, German, Italian, Japanese, Turkish, Dutch, Portuguese (PT & BR), Korean, Danish, Swedish, Polish, Norwegian, Hungarian, Greek, Finnish, Czech, Romanian, Croatian, Bulgarian, Slovak, Lithuanian, Estonian, Slovenian, Russian
Lock Function
Continuous / Auto-stabilization ("press to read") / At Interval
Measurement Method
Probe specific programmed method settings
Model
HQ2200 Multi/2 Channels
mV Measurement at Stable Reading
Yes
mV Measurement Range
Yes
mV Resolution
0.1 mV
Operating Error Messages
Clear text error messages displayed
Operating Humidity
90% (non-condensing)
Operating Interface
Soft Touch Keypad
Operating Temperature
0 - 60 °C
ORP Electrode Calibration
Predefined ORP standards (including Zobell's solution)
Parameter
pH
pH Buffer Sets
Colour-coded: 4.01, 7.00
10.01 pH; IUPAC
1.679, 4.005, 7.000, 10.012
12.45; DIN
1.09, 4.65, 9.23; User-defined custom buffer sets
pH Electrode calibration
1 - 3 Calibration points; Calibration summary data logged and displayed
pH Measurement Range
0 - 14 pH
pH Resolution
Selectable; 0.001/0.01/0.1 pH
Power Supply
Rechargeable lithium-ion battery 18650 (internal); Class II
USB power adapter
100 - 240 VAC, 50/60 Hz input; 5 VDC at 2 A USB power adapter output (external)
Probe Type
Intellical Standard Laboratory or Rugged Field
Probes included
PHC10105
Protection Class
IEC Class III (SELV (Separated/Safety Extra-Low Voltage) powered); external
Salinity Measurement Range
0 - 42 (ppt) (‰)
Salinity Resolution
0.01 (ppt) (‰)
Sensor
PHC10105; None
Sensor C
None
Storage Conditions
-20 - 60 °C, max. 90% relative humidity (non-condensing)
TDS Measurement Range
0.00 mg/L - 50.0 g/L NaCl
TDS Resolution
0.01 mg/L - 0.1 g/L upon measuring range
Temperature Compensation
Automatic Temperature compensation for pH
Temperature Measurement
°C or °F
Temperature Resolution
0.1 °C
Type of measurement
Field Measurements
Warranty
12 months
Weight
541 g (Meter only)
What's included?
LEV015.53.2200A HQ2200 Portable pH/ORP/EC/TDS/DO Multi-Meter, 2 channels; PHC10105 Intellical rugged field pH electrode, 5 m cable; LEZ015.99.A002A HQ rugged electrode field case; LEZ015.99A003A HQ hand strap & kickstand; LEZ015.99.A004A HQ protective glove; 2770020 Singlet single-use pH Buffer, pH 4.01, pk/20; 2770120 Singlet single-use pH Buffer, pH 7.00, pk/20; 2770220 Singlet single-use pH Buffer, pH 10.01, pk/20; 5828610 Cable marker for rugged electrode; User manuals

Standards & Compliance

HS Code
9027.89

Catalogue entry last verified: 2026-10-03. A manufacturer datasheet is attached to this entry — confirm application-critical values against it.

Frequently asked questions

Where can I buy LEV015.98.22009?
LEV015.98.22009 can be ordered from cballast.com as a Hach catalogue item: request a quotation on this page or email sales@cballast.com, and the order ships worldwide.
What is the manufacturer part number for this Hach product?
The manufacturer part number is LEV015.98.22009.
What is the HS code for LEV015.98.22009?
The customs tariff (HS) code for LEV015.98.22009 is 9027.89.
Who manufactures LEV015.98.22009?
LEV015.98.22009 is a Hach product in the Chemical Test Kits range.
Is a datasheet available for LEV015.98.22009?
Yes — the LEV015.98.22009 datasheet can be viewed and downloaded on this page.
What is the minimum order quantity for LEV015.98.22009?
The minimum order quantity is 1.
Is LEV015.98.22009 available?
Yes — LEV015.98.22009 is available to order: fulfilment is from supplier stock in cBallast's European network, dispatched directly from the warehouse to the customer, with the exact lead time stated in the quotation (reply within 24 hours).