PPCH Automated Pressure Controller / Calibrator

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
3072503  ·  HS 9031.80
PPCH Automated Pressure Controller / Calibrator

Fluke Calibration offers PPCH™ Automated Pressure ontroller/Calibrator for Hydraulic Pressure Applications.

Technical Specifications

Manufacturer part number (MPN)3072503
BrandFluke
CategoryTools
Power Requirements85 to 264 VAC, 50/60 Hz, 75 W max
Normal Operating Temperature Range15 to 35 °C
VibrationMeets MIL-T-28800D
Weight (Typical)50 kg approx (110 lb)
Dimensions30 cm H x 52 cm W x 50 cm D (12 in. x 20.5 in. x 20 in.) with enclosure 6U H rack mount version
Communications PortsRS232 (COM1, COM2), IEEE-488.2
Operating ModesGauge, absolute
Pressure RangesAtmosphere to 200 MPa (30,000 psi)
Operating MediaSebacate oil standard. Others optional (consult Fluke Calibration)
Internal Reservoir Volume250 cc (external unlimited)
Drive Air Supply70M - 500 kPa (75 psi), 300 l/m (10cfm) 140M - 500 kPa (75 psi), 450 l/m (15cfm) 100M - 700 kPa (100 psi), 300 l/m (10cfm) 200M - 700 kPa (100 psi), 450 l/m (15cfm)
Pressure ConnectionsDRIVE AIR SUPPLY 1/8 in. NPT F TEST DH500 (equivalent to AE F250C, HIP HF4)
Utility SensorPrecision/Resolution ± 0.10 % span / 0.001 % span
Drivers(8) 12V, 1 A max total output
CE ConformanceAvailable, must be specified
Control PrecisionTo ± 0.003 % of Q-RPT span
Control Volume0 to 100 cc, 50 cc optimum (operates in larger volumes but pressure stabilizing time increases)
Control Speed180 seconds into a 50 cc volume and 20 % FS excursions with default hold limits.
Lowest Controllable Pressure1 MPa (150 psi)(lower in optimum conditions and with PG7302)
Warm Up Time30 minute temperature stabilization recommended from cold power up
ResolutionTo 1 ppm, user adjustable
CalibrationA2LA accredited calibration report included
Q-RPTs Less than A200MA200M
Precision1 ± 0.012 % of reading5±0.015 % of reading5
Measurement Uncertainty Delivered ± 0.013 % of reading5± 0.018 % of reading5
± 0.016 % of reading5± 0.020 % of reading5

Properties

MOQ1

Cross-References