| Accuracy | ± 5% of reading ± 0.1 mg/L NO₃-N |
| Altitude | 2000 m (6562 ft) maximum |
| Ambient Temperature | 2 - 40 °C (36 - 100 °F), 95% relative humidity, non-condensing |
| Cable Length | 10 m (33 ft); Extension cables are available: 5, 10, 15, 20, 30 and 50 m. The maximum cable length is 60 m (190 ft). |
| Certifications | CE, CMIM, UKCA, FCC, and ISED approved |
| Controller Compatibility | SC4500, SC1000, SC200 |
| Dimensions | 70 x 402 mm (3 x 15.8 inches) approximately |
| Indoor/Outdoor | Outdoor |
| IP Rating | IP68 |
| Lower Limit of Detection (LOD) | 0.1 mg/L NO₃-N |
| Measurement Method | UV Absorption, sludge compensated, 2-channel beam path |
| Measuring Interval | 15, 30 seconds, 1, 5, 10, 30 minutes |
| Measuring Principle | UV absorption measurement, reagent-free |
| Model | NT3100sc |
| Over voltage category | II |
| Parameter | Nitrate |
| Path Length | 1 mm |
| Pollution Degree | 2 |
| Power Consumption | 9 |
| Pressure Range | Sensor pressure limit: 0.5 bar |
| Process Connection | Immersion directly in media; Bypass with Flow Through Unit; Sedimenter |
| Range | 0.1 - 90 mg/L NO₃-N |
| Resolution | 0.01 - 999.9 |
| Response Time | T100: 1 minute |
| Sample Temperature | 2 - 40 °C (36 - 100 °F) |
| Signal Average Time | 1 to 12 measurements |
| Sludge Compensation | Yes |
| Units | mg/L, ppm |
| Warranty | 12 months |
| Weight | 4.8 kg (10.58 lb) with 10 m cable |
| What's included | Sensor with cable set, set of wipers, user manual |