| Manufacturer part number (MPN) | 5282560 |
| Brand | Fluke |
| Category | Tools |
| Measurement Function | Specification |
| Full scale reading | 999 counts |
| Accuracy at 4 s to 10 us/div | ± (1.5 % + 6 counts) |
| DC to 60 Hz | ± (1.5 % + 10 counts) |
| 60 Hz to 20 kHz | ± (2.5 % + 15 counts) |
| 20 kHz to 1 MHz | ± (5 % + 20 counts) |
| 1 MHz to 25 MHz | ± (10 % + 20 counts) |
| Purpose | To measure on pulse width modulated signals, like motor drive inverter outputs |
| Principle | Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency |
| Accuracy | As Vac+dc for sinewave signals |
| Modes | Max peak, min peak, or pk-to-pk |
| Max peak, min peak | ± 0.2 division |
| Pk-to-pk | ± 0.4 division |
| Ranges | Same as V ac, Vac+dc or V peak |
| Scale Factors | 0.1 mV/A, 1 mV/A, 10 mV/A, 20 mV/A, 50mV/A, 100 mV/A, 200 mV/A, 400 mV/A |
| Range | 1.000 Hz to 500 MHz |
| Readings | Voltage difference (dV), time difference (dt), voltage vs time difference (dV/dt), overshoot |
| Harmonics | DC to 51st |
| Spectrum ranges | 1…9 kHz, 9-150 kHz (20 MHz filter on), up to 500 MHz (voltage modulation) |
| Events / second | Indicative percentage based on rise and fall time (Impulse discharges) measurements |
| Number of screens | Typical 50 screens can be saved in reports (depends on compression ratio) |
| Transfer to PC | Using 32 GB or smaller 2 GB USB stick or mini-USB to USB cable or WiFi link and FlukeView™ 2 for ScopeMeter® |
| Voltage Probe | 1:1, 10:1, 100:1, 1000:1, 20:1, 200:1 |
| Current Clamp | 0.1 mV/A, 1 mV/A, 10 mV/A, 20 mV/A, 50 mV/A, 100 mV/A, 200 mV/A, 400 mV/A |
| Shaft Voltage Probe | 1:1, 10:1, 100:1 |
| General | IEC 61010-1: Pollution Degree 2 |
| Measurement | Measurement IEC 61010-2-030: CAT IV 600 V / CAT III 1000 V |
| Maximum voltage between any Terminal and Earth Ground | 1000 V |
| Max. input voltages | Via VPS410-II or VPS421 1000 V CAT III / 600 V CAT IV |
| BNC Input | A, B, C, D directly 300 V CAT IV |
| Working voltage between probe tip and probe reference lead | VPS410-II: 1000 V VPS421: 2000 V |