9172 Metrology Wells. Now we’ve combined bath-level performance with dry-well functionality and legitimate reference thermometry to create Metrology Wells. Metrology Wells light you bring lab-quality performance into whatever field environment you might work in.

Technical Specifications

Manufacturer part number (MPN)2459286
BrandFluke
CategoryTools
Range (at 23°C ambient)35°C to 425°C (95°F to 797°F)
Display accuracy±0.1°C: 35°C to 100°C ±0.15°C: 100°C to 225°C ±0.2°C: 225°C to 425°C
Stability±0.005°C: 35°C to 100°C ±0.008°C: 100°C to 225°C ±0.01°C: 225°C to 425°C
Axial uniformity±0.05°C: 35°C to 100°C ±0.1°C: 100°C to 225°C ±0.2°C: 225°C to 425°C
Radial uniformity±0.01°C: 35°C to 100°C ±0.02°C: 100°C to 225°C ±0.025°C: 225°C to 425°C
Hysteresis0.04°C
Well depth203 mm (8 in)
Resolution0.001°C
DisplayLCD, °C or °F, user-selectable
Key padTen key with decimal and ± button. Function keys, menu key, and °C / °F key.
Cooling time220 min: 425°C to 35°C 100 min: 425°C to 100°C
Heating time27 min: 35°C to 425°C
Size (H x W x D)366 x 203 x 323 mm (14.4 x 8 x 12.7 in)
Weight13.2 kg (29 lb)
Power115 V AC (±10%), or 230 V AC (±10%), 50/60 Hz, 1025 W
Computer interfaceRS-232 Interface
Traceable calibration (NIST)Data at 100°C, 150°C, 250°C, 350°C, and 425°C
SpecificationsBuilt-in Reference Input
Temperature range–200°C to 962°C (–328°F to 1764°F)
Resistance range0 Ω to 400 Ω, auto-ranging
CharacterizationsITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d
Resistance accuracy0 Ω to 20 Ω: 0.0005 W 20 Ω to 400 Ω: 25 ppm
Resistance resolution0 Ω to 20 Ω: 0.0001 Ω 20 Ω to 400 Ω: 0.001 Ω
Measurement period1 second
Probe connection4-wire with shield, 5-pin DIN connector
CalibrationNVLAP accredited (built-in reference input only), NIST-traceable calibration provided

Properties

MOQ1

Cross-References