| Manufacturer part number (MPN) | 5319188 |
| Brand | Fluke |
| Category | Tools |
| Functions | ACR, DCV, enclosure potential measurement |
| Measurement method | AC 4-wire method (1 kHz) |
| Rated input | DC ±11V |
| Input impedance | 10 MΩ, high-z (>10 GΩ) |
| Sampling speed | Ex-Fast, Fast, Medium, Slow |
| Allowable total line resistance | ≤ 10Ω @ 3 mΩ measurement range |
| Trigger | Internal/external |
| Trigger delay | ON/OFF |
| Delay time | 0 to 9.999 s |
| Average | ON/OFF |
| Average count | 2 to 16 |
| Measurement storage | Up to 512 sets |
| Channel extension | Internal: Up to 2 SW9010 multiplexer cards can be added for up to 64 channels (BT5311 and BT5321 only) External: Up to 256 channels by using SW1080 switch mainframe |
| External interface | RS-232C , LAN |
| Other functions | Contact check |
| Ex-Fast | 10 ms / 8.3 ms |
| Fast | 20 ms / 16.7 ms |
| Medium | 100 ms / 83.3 ms |
| Slow | 200 ms / 166.7 ms |
| Range 3 mΩ | 30 mΩ |
| Max. Display value 5.0000 mΩ | 50.000 mΩ |
| Resolution 0.1 µΩ | 1 µΩ |
| Test current1 100/200/300 mA² | 100 mA |
| Test current frequency | 1 kHz ± 1Hz |
| Accuracy³ | 0.2% * rdg + 6 dgt |
| Temperature coefficient ( 28 °C) | (0.05% * rdg + 0.005% * F.S.) / °C |
| BT5310/BT5311 | BT5320/BT5321 |
| Digit 6.5 digits | 7.5 digits |
| Range | 10 V |
| Max. Display value ±11.00000V | ±11.000000V |
| Resolution 10 µV | 1 µV |
| Accuracy1 25 ppm * rdg + 50 µV | 18 ppm * rdg + 25 µV |
| (5 ppm * rdg + 1 ppm * F.S.) / °C | (1 ppm * rdg + 1 ppm * F.S.) / °C |
| Features BT5310 | BT5311 |
| DCV resolution 6.5 digit | 7.5 digit |
| No | Yes. Up to 2 cards |
| Dimensions (W x H x L) | BT5300 Battery Tester: 216 mm x 133 mm x 383 mm |
| Weight | BT5300 Battery Tester: 7.3 kg (Without SW9010) |
| Display | 4.3-inch TFT LCD, resolution 480 x 272 |
| Operation temperature | 0 °C to +40 °C |
| Storage temperature | -10 °C to +50 °C |
| Altitude | 3000 m |
| Storage altitude | 12000 m |
| Rated voltage | AC 100V to 240V (Auto-selection) |
| Rated frequency | 50/60Hz |
| Rated power | 60 VA |
| Safety | IEC 61010-1: Pollution Degree 2 |
| EMC | IEC 61326-1: Industrial |
| Warranty | BT5300 Battery Tester: 3 years SW1080 Switch Mainframe: 3 years SW9010 Multiplexer Card: 1 year Note: Relays, fuses and connectors are not covered by warranty. |