cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-08
sales@cballast.com
cballast.com

Fluke 729 Pro Automatic Pressure Calibrator

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
5300476  ·  HS 9031.80
Fluke 729 Pro Automatic Pressure Calibrator

Automatic pressure generation and regulation up to 1000psi, 7Mpa with plug-in replaceable pressure control modules, Fluke 729Pro not only simplifies pressure calibraiton but saves time,reduces errors and improves accuracy

Technical Specifications

Manufacturer part number (MPN)5300476
BrandFluke
CategoryTools
Key parameters 729 Pro 7M729 Pro 4M
-13 to 1000 psi, -90 to 7000 kPa, -0.90 to 70 bar-13 to 600 psi, -90 to 4000 kPa, -0.90 to 40 bar
Pressure typeGauge/absolute (Depends on the type of internal or external pressure module
Pressure accuracy and Control stability0.02% of FS, 0.005% of FS
Replaceable pressure measurement modules Yes, 6 modulesYes, 4 modules
mA sourcing/measurementYes
Voltage measurementYes
Loop power24 V dc
Pressure switch testingYes
Pressure leakage testingYes
HART CommunicationYes, basic commands supported
Autostep / auto rampYes (with customizable speed, mode and style)
BatteryRechargeable lithium battery with charge indication (charged in the instrument or separately)
LanguageEnglish
mA measure, source and simulate0 mA to 24 mA
Volts dc measurement0 V dc to 30 V dc
mA dc source, simulate and measure1 μA
Voltage dc measurement1 mV
Accuracy0.01% FS ±2 LSD all ranges (at 23 °C ±5 °C)
Stability20 ppm FS /°C (-10 °C to +18 °C and 28 °C to 50 °C)
mA simulation working voltage12 V dc to 30 V dc
Loop compliance voltage24 V dc @ 20 mA
Temperature measurement , 100 Ω Pt(385) RTD-50 °C to 150 °C
Temperature resolution0.01 °C
Temperature accuracy±0.1 °C ±0.25 °C combined uncertainty when using 720 RTD probe (optional accessory)
Drive capability (mA source)1200 Ω (without HART resistor), 950 Ω (with internal HART resistor)
Dimension (HxWxL)290 x 215 x100 mm, (4.3 x 4.4 x 9.4 in)
Weight4.9 kg, (10.8 lbs)
Operating temperature-10 °C to 50 °C
Pressure control0 °C to 50 °C
Battery charging0 °C to 40 °C
Storage temperature-20 °C to 60 °C
Operating altitude<3000 m
Storage altitude<13000 m
Operating humidityNon-condensing (<10°C) ≤90 % RH (10 °C to 30 °C) ≤75 % RH (30 °C to 40 °C) ≤45 % RH (40 °C to 50 °C)
Pressure module Pressure typeRange
FLK-PMM-200K Gauge(-15 to 30) psi
FLK-PMM-200K (-100 to 200) kPa0.001 kPa
FLK-PMM-200K (-1 to 2) bar0.00001 bar
FLK-PMM-200KA Absolute(0 to 30) psi
FLK-PMM-200KA (0 to 200) kPa0.001 kPa
FLK-PMM-200KA (0 to 2) bar0.00001 bar
FLK-PMM-2000K Gauge(-14 to 300) psi
FLK-PMM-2000K (-95 to 2000) kPa0.01 kPa
FLK-PMM-2000K (-0.95 to 20) bar0.0001 bar
FLK-PMM-4000K Gauge(-14 to 600) psi
FLK-PMM-4000K (-95 to 4000) kPa0.01 kPa
FLK-PMM-4000K (-0.95 to 40) bar0.0001 ba
FLK-PMM-7000K Gauge(-14 to 1000) psi
FLK-PMM-7000K (-95 to 7000) kPa0.01 kPa
FLK-PMM-7000K (-0.95 to 70) bar0.0001 ba
FLK-PMM-7000KA Absolute(0 to 1000) psi
FLK-PMM-7000KA (0 to 7000) kPa0.01 kPa
FLK-PMM-7000KA (0 to 70) bar0.0001 ba

Properties

MOQ1

Cross-References

Attached Documents