| Manufacturer part number (MPN) | 1617262 |
| Brand | Fluke |
| Category | Tools |
| Current measurement Range | 0 to 24 mA |
| Current measurement Resolution | 0.001 mA |
| Current measurement Accuracy | 0.015% Rdg + 2 LSD |
| Current sourcing Range | 0 to 20 mA or 4 to 20 mA |
| Current sourcing Accuracy | 0.015% Rdg + 2 LSD |
| Current sourcing Drive capability | 1200 Ω 950 Ω at 20 mA in HART mode |
| Current simulation Range | 0 to 20 mA or 4 to 20 mA |
| Current simulation Accuracy | 0.015% Rdg + 2 LSD |
| Current simulation Ext. loop voltage requirement | 12-30 V |
| Loop power | While measuring mA, 24 V |
| Voltage measurement Range | 0 to 28 V DC |
| Voltage measurement Resolution | 1 mV |
| Voltage measurement Accuracy | 0.015% Rdg + 2 LSD |
| Operating temperature | -10°C to 55°C |
| Storage temperature | -30°C to 60°C |
| Humidity (without condensation) | 0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C) |
| Temperature coefficient | -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C |
| Safety | CSA C22.2 No. 1010.1:1992 |
| EMC | EN50082-1:1992 and EN55022:1994 Class B |
| Maximum voltage | 30 V |
| Size | 152 x 70 x 41 mm |
| Weight | 224 g (340 g including holster) |
| Batteries | One 9 V battery Alkaline |
| Warranty | Three years |
| Shock and vibration | Per MIL-T-28800 for a Class 2 Instrument |
| Display current and % of span | Yes |
| Auto step, auto ramp | Yes |
| Span check | Yes |