| Manufacturer part number (MPN) | 4489996 |
| Brand | Fluke |
| Category | Tools |
| Voltage | Resistance |
| > Voltage Lower < Voltage Lower | < Reference |
| Pass Fail | Pass |
| Range | Resolution |
| Battery Resistance / Strap Resistance1 3 mΩ | 0.001 mΩ |
| 30 mΩ 0.01 mΩ | 0.8% + 6 |
| 300 mΩ 0.1 mΩ | 0.8% + 6 |
| 3000 mΩ 1 mΩ | 0.8% + 6 |
| VDC 6 V | 0.001 V |
| 60 V 0.01 V | 0.09% + 5 |
| 600 V 0.1 V | 0.09% + 5 |
| 1000 V 1 V | 0.09% + 5 |
| VAC (45 Hz to 500 Hz with 800 Hz filter) 600 V | 0.1 V |
| Frequency (displayed with VAC and AAC)2 500 Hz | 0.1 Hz |
| AC voltage ripple (20 KHz Max) 600 mV | 0.1 mV |
| 6000 mV 1 mV | 3% + 10 |
| ADC/AAC (with accessory Fluke i410) 400 A | 1 A |
| Temperature 0°C to 60°C | 1°C |
| Meter mode | 999 records for each measurement position with time stamp |
| Sequence mode | Up to 100 profiles and 100 profile templates (Each profile stores up to 450 batteries) with time stamp |
| Measurement Modes BT510 | BT520 |
| Size (H x W x D) | 22 x 10.3 x 5.8 cm (9 x 4 x 2 in) |
| Weight | 850 g (1.9 lb) |
| Screen dimensions | 7.7 x 5.6 cm (3 x 2.2 in) |
| Interface | USB mini |
| Warranty | 3 years |
| Long angled probe (overall length) | 242.2 cm (95.38 in) |
| Short angled probe (overall length) | 216.8 cm (85.38 in) |
| Overall span (coil relaxed) | 10.1 cm (4 in) |
| Overall maximum span (coil extended) | 109.2 cm (43 in) |
| BTL20ANG probe tip angle | 20 degrees from horizontal |
| Operating temperature | 0°C to 40°C |
| Storage temperature | -20°C to 50°C |
| Lithium-ion battery charging temperature | 0°C to 40°C |
| Operating humidity | Non-condensing (10°C) |
| Operating altitude | Sea level to 2000 meters |
| Storage altitude | Sea level to 12,000 meters |
| IP rating | IP40 |
| Radio | FCC Class A |
| Vibration requirements | MIL-PRF-28800F: Class 2 |
| Drop test requirements | 1 meter |
| Temperature coefficients | Add 0.1 x specified accuracy for each degree C above 28°C or below 18°C |
| Safety compliance | 600 V CAT III |
| EMC | IEC 61326 |
| ROHS | China, Europe |
| Protection class 2 | Pollution Degree II |
| Battery compliance | UN38.3 |