cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

Fluke 113 Digital Multimeter

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
3078875  ·  HS 9030.33
Fluke 113 Digital Multimeter

The Fluke 113 is an easy-to-use basic multimeter with VCHEK™ LoZlow impedance measurement function for ghost voltage detection which can measure and display both AC and DC voltages.

Technical Specifications

Manufacturer part number (MPN)3078875
BrandFluke
CategoryTools
VCHEK1 2 3 Range6.000 V / 60.00 V / 600.0 V
VCHEK1 2 3 Resolution0.001 V / 0.01 V/ 0.1 V
VCHEK1 2 3 AccuracyDC, 45 to 500 Hz: 2.0% + 3 500 Hz to 1 kHz: 4.0% + 3
Diode test4 Range2.000 V
Diode test4 Resolution0.001 V
Diode test4 Accuracy2.0% + 3
Continuity4 AccuracyBeeper on 250 Ω detects opens or shorts of 500 μs or longer
Ω4 Range600.0 Ω / 6.000 kΩ / 60.00 kΩ
Ω4 Resolution0.1 Ω / 0.001 kΩ / 0.01 kΩ
Ω4 Accuracy0.9% + 2 / 0.9% + 1 / 0.9% + 1
Capacitance4 Range1000 nF / 10.00 μF / 100.0 μF / 9999 μF
Capacitance4 Resolution1 nF / 0.01 μF / 0.1 μF / 1 μF
Capacitance4 Accuracy1.9% + 2 100 μF - 1000 μF: 1.9% + 2 1000 μF: 5% + 20
Maximum voltage between any terminal and earth ground600 V
Display3 3/4-digits, 6000 counts, updates 4/sec
Operating temperature-10°C to 50°C (14°F to 122°F)
Storage temperature-40°C to 60°C (-22°F to 140°F)
Temperature coefficient0.1 x (specified accuracy) / °C ( 28°C)
Operating altitude2,000 m
Storage altitude10,000 m
Relative humidity95% to 30°C 75% to 40°C 45% to 50°C
Battery type9 Volt Alkaline, IEC 6F22
Battery lifeAlkaline: 300 hours typical, without backlight
Size (H x W x L)167.1 x 85.1 x 46.0 mm (6.58 x 3.35 x 1.81 in)
Weight13.0 oz (404 g)
SafetyIEC 61010-1: Pollution Degree 2 IEC 61010-2-033: CAT IV 600V
EMI regulationsComplies with FCC Part 15, Class B
VCHEK Input impedance (nominal)~3 kΩ < 300 pF
VCHEK Common mode rejection ratio (1 kΩ unbalanced)> 60 dB at DC, 50 or 60 Hz
Ω Open circuit test voltage< 2.7 VDC
Ω Full scale voltage< 0.7 VDC
Diode test Open circuit test voltage2.7 VDC
Diode test Full scale voltage2.000 VDC
Ω Short circuit current< 350 μA
Diode test Short circuit current< 1.0 mA

Properties