Fluke 707 is a milliamp loop calibrator sourcing and measuring 0-24 mA at 0.001 mA resolution with 0.015% accuracy, and measuring 0-28 V DC. Provides 24 V loop power, drives 1200 Ohm at 20 mA and includes a 250 Ohm HART resistor for compatibility with HART transmitters. Used by instrumentation technicians for 4-20 mA loop commissioning, transmitter simulation and control-valve checks in process plants.
| Manufacturer part number (MPN) | 14579 |
| EAN / GTIN | 0095969107150 |
| Sales unit | pcs |
| Manufacturer model | Fluke 707 |
| Current measurement | Range 0 to 24 mA |
| Resolution | 0.001 mA |
| Accuracy | 0.015% Rdg + 2 LSD |
| Current sourcing | Range 0 to 20 mA or 4 to 20 mA |
| Drive capability | 1200 Ω 950 Ω at 20 mA in HART mode |
| Current simulation | Range 0 to 20 mA or 4 to 20 mA |
| Ext. loop voltage requirement | 12-30 V |
| Loop power | While measuring mA, 24 V |
| Voltage measurement | Range 0 to 28 V DC |
| Operating temperature | -10°C to 55°C |
| Storage temperature | -30°C to 60°C |
| Humidity (without condensation) | 0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C) |
| Temperature coefficient | -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C |
| Safety | CSA C22.2 No. 1010.1:1992 |
| EMC | EN50082-1:1992 and EN55022:1994 Class B |
| MOQ | 1 |