cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

TU5300 sc Turbidimeter — with automatic cleaning (ACM)

Hach · Chemical Test Kits · Water Analysis Meters · USA · hach.com
Part Number
LXV445.99.30122  ·  HS 9027.89
TU5300 sc Turbidimeter — with automatic cleaning (ACM)

Hach TU5300 sc Turbidimeter — with automatic cleaning (ACM). Reference part number LXV445.99.30122.

Technical Specifications

Accuracy±2% or 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration MethodFor Formazin and Stablcal®; 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range; Custom calibration for up to 6-point calibrations
CertificationsCE compliant; US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version; Australian ACMA Marking
ControllerSensor Only
Controller CompatibilitySC4500, SC1000, SC200
Dimensions (H x W x D)249 mm x 268 mm x 190 mm
Enclosure RatingElectronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting TypeSample quick connector: ¼-in. for ¼-in. tubing
Light SourceClass 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Lower Limit of Detection (LOD)0.002 NTU
MaterialASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40; Thermoplastic Elastomer TPS-SEBS
ModelTU5300sc
Mounting ConfigurationsWall Mount
Operating HumidityRelative humidity: 5 - 95% at different temperatures, non-condensing
Operating Temperature Range0 - 50 °C (32 - 122 °F)
OptionsProcess Head with Automatic Cleaning Module
RangeISO; 0 - 1000 NTU / FNU / TE/F / FTU; 0 - 250 EBC
RegulatoryISO
RepeatabilityBetter than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution0.0001 NTU / FNU / TE/F / FTU / EBC
Response TimeT90 <45 seconds at 100 mL/min
Sample Flow Rate100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature2 - 60 °C (35 - 140 °F)
Signal Average Time30 - 90 seconds
Storage Conditions-40 - 60 °C (-40 - 140 °F)
Stray Light<10 mNTU
UnitsNTU, FNU, TE/F, FTU, EBC
VerificationLiquid: Stablcal®
Formazin (0.1 - 40 NTU); DryGlass Rod at <0.1 NTU
Warranty12 months
Weight5.95 lbs. (2.7 kg); 11 lbs. (5.0 kg) with all accessories
What's includedTU5300sc Turbidimeter, Automatic Cleaning Unit, User Manual, Wall Mount, Vial Replacement Tool, Screw Set, Drying Cartridge, Flow Regulator, Service Bracket

Properties

MOQ1

Approvals

Attached Documents