| Communication interface | RFID |
| Accuracy | ±2% plus 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000 NTU based on Formazin primary standard (at 25 °C) |
| Calibration Method | For Formazin and Stablcal; 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range; Custom calibration for up to 6-point calibrations; Degrees 0 - 100 mg/L; SDVB 0 - 1000 NTU |
| Certifications | CE compliant; US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version; Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50); Australian ACMA Marking |
| Communication | Link2sc |
| Dimensions (H x W x D) | 195 mm x 409 mm x 278 mm |
| Display | 7" WVGA (800 x 480) color TFT-LCD with touch screen |
| Enclosure Construction | PC/ABS Blend Plastic |
| Light Source | Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50) |
| Model | TU5200 ISO |
| Operating Humidity | 80% at 30 °C (non condensing) |
| Operating Temperature Range | 10 - 40 °C (50 - 104 °F) |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 100 - 240 VAC |
| Range | EPA; 0 - 700 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 175 EBC; ISO; 0 - 1000 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 250 EBC |
| Regulatory | DIN EN ISO 7027 |
| Repeatability | <40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater; >40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F) |
| Resolution | 0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L |
| Response Time | Signal averaging off: 7 seconds; Signal averaging on: 10 seconds (when averaging time is 5 seconds) |
| Sample Temperature | 4 - 70 °C (39 - 158 °F) |
| Source Lamp | ISO Version - 850 nm Class 2 Laser |
| Storage Conditions | -30 - 60 °C (-22 - 140 °F) |
| Stray Light | <10 mNTU |
| Units | NTU, FNU, TE/F, FTU, EBC; mg/L if calibrated with Degrees calibration curve |
| Warranty | 12 months |
| Weight | 2.4 kg (5.29 lbs.) |
| What's included | TU5200 benchtop turbidimeter, user manual, Stablcal kit (10, 20 and 600 NTU), 2 sample vials, power supply, dust cover, vial stand |