cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

PG7302 Piston Gauge

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
3069747
PG7302 Piston Gauge

Fluke Calibration's PG7302™ Piston Gauge line features reference level pressure standards offering a unified solution from vacuum to 500 MPa.

Technical Specifications

Manufacturer part number (MPN)3069747
BrandFluke
CategoryTools
Power requirements85 V ac to 264 V ac, 47 Hz to 440 Hz, 22 VA max consumption
Normal operating temperature range18 °C to 28 °C (64 °F to 82 °F)
WeightInstrument platform with no mass loaded: PG7302 13 kg (28 lb) PG terminal 1.4 kg (3 lb)
DimensionsInstrument platform (HxWxD): 36 cm x 40 cm x 35 cm (14.5 in x 15.8 in x 13.8 in) Height: Top of mounting post with piston-cylinder module installed PG terminal (HxWxD): 12 cm x 15 cm x 20 cm (4.7 in x 5.9 in x 7.9 in)
Communications portsRS-232 COM1: Host computer COM2: AMH automated mass handler, external barometer COM3: Automated pressure controller IEEE-488.2: Host computer
Maximum pressure rangesActual range depends on piston-cylinder and mass set selection.
Operating mediaOil: ethyl hexyl Sebacate (oil)
Maximum mass load100 kg
Pressure connectionsTest port: DH500 Note: DH200 and DH500 are gland and collar type fittings for 1/4 in (6 mm) coned and left hand threaded tube. DH200 is equivalent to AE SF250C, HIP LF4, etc. DH500 is equivalent to AE F250C, HIP HF4, etc.
CE conformanceAvailable, must be specified
Mass set specificationsAll masses are delivered in molded, reusable transit cases with custom inserts.
MassesMaterial: 304L non-magnetic stainless steel Finish: Electropolished Adjustment tolerance: ± 20 ppm of nominal value Accuracy of measured values: ± 5 ppm or 1 mg, whichever is greater Traceability of measured values: National Institute of Standards and Technology (NIST)
Piston-Cylinder TemperatureRange: 0 to 40 degrees C Resolution: 0.01 degrees C Accuracy: ± 0.1 degrees C
Ambient TemperatureRange: 0 to 40 degrees C Resolution: 0.1 degrees C Accuracy: ± 1 degrees C
Barometric pressure with internal sensorRange: 70 kPa to 110 kPa Resolution: 10 Pa Accuracy: ± 140 Pa Barometric pressure can also be read automatically from any RS-232 device such as RPM4
Relative humidityRange: 5 % to 95 % RH Resolution: 1 % RH Accuracy: ± 10 % RH
Piston positionRange: ± 4.5 mm Resolution: 0.1 mm
Piston rotation (rate and deceleration)Range: 2 to 150 rpm Resolution: 1 rpm

Properties

MOQ1

Cross-References

Attached Documents