cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

MOLBOX2 Flow Terminal

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
6074904
MOLBOX2 Flow Terminal

The molbox2 Flow Terminal enables you to achieve the lowest uncertainty available for gas flow meter and controller calibrations.

Technical Specifications

Manufacturer part number (MPN)6074904
BrandFluke
CategoryTools
Usable range with SPcalibration and molbox2A700KUsable range withHP calibration andmolbox2-S -A2M
1E2-S 15 to 60 slm*20 to 300 slm
5E2-S 67 to 300 slm*100 to 1000 slm
Power requirements85 V ac to 264 V ac, 47 Hz to 440 Hz, 18 VA max consumption
Normal operating temperature range15 °C to 30 °C (59 °F to 86 °F)
Storage temperature range-20 °C to 70 °C (-4 °F to 158 °F)
VibrationMeets MIL-T-28800D
Weight6.8 kg (15 lb) max
Dimensions (WxHxD)32 cm x 12 cm x 30 cm (12.6 in x 4.7 in x 11.8 in) approx.
Communication portsRS-232 (COM1), RS-232 (COM2), IEEE-488.2
Pressure connections (molbox2 and molbloc)Quick connectors equivalent to Swagelok® QM Series (SS-QM2-B200)
Flow ranges5000 slm. See separate molbloc-L and molbloc-S range tables
Flow measurement rate340 milliseconds
Valve driver option(8) 12 V outputs. Each output can sink 500 mA at 12 V, max 1 Amp total
MFC control option (analog input/output)Nominal voltage range: 0 V dc to 6 V dc input, 0 V dc to 5 V dc output Nominal current range: 4 mA to 20 mA input, 4.01 mA to 20 mA output Accuracy: ± 0.1 % FS (set), ± 0.05 % FS (measure)
TypeQ-RPT Characterized Quartz Reference Pressure Transducers – Oscillating quartz crystal with mechanical bellows
A700K0 to 600 kPa absolute (0 to 87 psia)
A350K0 to 300 kPa absolute (0 to 44 psia)
S A1.4M (molbloc-S only)0 to 1.2 MPa absolute (0 to 174 psia)
S A2M (molbloc-S only)0 to 2 MPa absolute (0 to 290 psia)
Absolute pressure⁴± (0.01 % of reading or 0.003 % Q-RPT span, whichever is greater)
Differential pressure (A700K with Tare)± (8.4 Pa (0.0012 psi) or 0.032 % ΔP, whichever is greater)
Differential pressure (A350K with Tare)± (4.2 Pa (0.0006 psi) or 0.026 % ΔP, whichever is greater)
Range (FS)0 to 40 °C
Resolution0.01 °C
molbloc PRT precision± 0.02 °C (15 to 30 ˚C)
Reference resistor100 and 110 W ± 0.01 %, stability < 25 ppm/year
Ohmic measurement± 0.02 % of reading (15 °C to 30 °C)
Standard molbloc calibrationPremium molbloc calibration
± 0.09 % of reading, ± 0.009 % FS under 10 % FS± 0.03 % of reading, ± 0.003 % FS under 10 % FS
± 0.2 % of reading, ± 0.02 % FS under 10 % FS± 0.125 % of reading, ± 0.0125 % FS under 10 % FS
Stability (one-year)² ± 0.05 % of reading± 0.03 % of reading
± 0.2 % of reading± 0.125 % of reading

Properties

MOQ1

Cross-References

Attached Documents