| Manufacturer part number (MPN) | 6074904 |
| Brand | Fluke |
| Category | Tools |
| Usable range with SPcalibration and molbox2A700K | Usable range withHP calibration andmolbox2-S -A2M |
| 1E2-S 15 to 60 slm* | 20 to 300 slm |
| 5E2-S 67 to 300 slm* | 100 to 1000 slm |
| Power requirements | 85 V ac to 264 V ac, 47 Hz to 440 Hz, 18 VA max consumption |
| Normal operating temperature range | 15 °C to 30 °C (59 °F to 86 °F) |
| Storage temperature range | -20 °C to 70 °C (-4 °F to 158 °F) |
| Vibration | Meets MIL-T-28800D |
| Weight | 6.8 kg (15 lb) max |
| Dimensions (WxHxD) | 32 cm x 12 cm x 30 cm (12.6 in x 4.7 in x 11.8 in) approx. |
| Communication ports | RS-232 (COM1), RS-232 (COM2), IEEE-488.2 |
| Pressure connections (molbox2 and molbloc) | Quick connectors equivalent to Swagelok® QM Series (SS-QM2-B200) |
| Flow ranges | 5000 slm. See separate molbloc-L and molbloc-S range tables |
| Flow measurement rate | 340 milliseconds |
| Valve driver option | (8) 12 V outputs. Each output can sink 500 mA at 12 V, max 1 Amp total |
| MFC control option (analog input/output) | Nominal voltage range: 0 V dc to 6 V dc input, 0 V dc to 5 V dc output Nominal current range: 4 mA to 20 mA input, 4.01 mA to 20 mA output Accuracy: ± 0.1 % FS (set), ± 0.05 % FS (measure) |
| Type | Q-RPT Characterized Quartz Reference Pressure Transducers – Oscillating quartz crystal with mechanical bellows |
| A700K | 0 to 600 kPa absolute (0 to 87 psia) |
| A350K | 0 to 300 kPa absolute (0 to 44 psia) |
| S A1.4M (molbloc-S only) | 0 to 1.2 MPa absolute (0 to 174 psia) |
| S A2M (molbloc-S only) | 0 to 2 MPa absolute (0 to 290 psia) |
| Absolute pressure⁴ | ± (0.01 % of reading or 0.003 % Q-RPT span, whichever is greater) |
| Differential pressure (A700K with Tare) | ± (8.4 Pa (0.0012 psi) or 0.032 % ΔP, whichever is greater) |
| Differential pressure (A350K with Tare) | ± (4.2 Pa (0.0006 psi) or 0.026 % ΔP, whichever is greater) |
| Range (FS) | 0 to 40 °C |
| Resolution | 0.01 °C |
| molbloc PRT precision | ± 0.02 °C (15 to 30 ˚C) |
| Reference resistor | 100 and 110 W ± 0.01 %, stability < 25 ppm/year |
| Ohmic measurement | ± 0.02 % of reading (15 °C to 30 °C) |
| Standard molbloc calibration | Premium molbloc calibration |
| ± 0.09 % of reading, ± 0.009 % FS under 10 % FS | ± 0.03 % of reading, ± 0.003 % FS under 10 % FS |
| ± 0.2 % of reading, ± 0.02 % FS under 10 % FS | ± 0.125 % of reading, ± 0.0125 % FS under 10 % FS |
| Stability (one-year)² ± 0.05 % of reading | ± 0.03 % of reading |
| ± 0.2 % of reading | ± 0.125 % of reading |