Fluke 9190A Ultra-Cool Field Metrology Well

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
4178780  ·  HS 9031.80
Fluke 9190A Ultra-Cool Field Metrology Well

The Fluke 9190A Ultra-Cool Field Metrology Well is the most accurate and stable, cold temperature dry-block on the market.

Technical Specifications

Manufacturer part number (MPN)4178780
BrandFluke
CategoryTools
Temperature range at 23°C–95°C to 140°C (–139°F to 284°F)
Display accuracy±0.2°C full range
Accuracy with external reference3±0.05°C full range
Stability±0.015°C full range
Axial uniformity at 40 mm (1.6 in)±0.05°C full range
Radial gradient±0.01°C full range
Loading effect(with a 6.35 mm reference probe and three 6.35 mm probes)
Operating conditions0°C to 35°C, 0% to 90%
Immersion (well) depth160 mm (6.3 in)
Well diameter30 mm (1.18 in)
Heating time1–95°C to 140°C: 40 min
Cooling time123°C to –90°C: 80 min
Stabilization time215 min
Resolution0.01°
DisplayLCD, °C or °F user selectable
Size (H x W x D)480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
Weight16 kg (35 lb)
Power requirements100 V to 115 V (±10%) 50/60 Hz, 575 W
System fuse ratings115 V: 6.3 A T 250 V
4–20 mA fuse (-P model only)50 mA F 250 V
Computer interfaceRS-232, and USB Serial included
SafetyIEC 61010-1, Installation Category II, Pollution degree 2
Electromagnetic environmentIEC 61326-1: Basic
RefrigerantsR32 (Difluoromethane)
Reference resistance range0 Ω to 400 Ω
Reference resistance accuracy40 Ω to 42 Ω: ±0.0025 Ω 42 Ω to 400 Ω: ±60 ppm of reading
Reference characterizationsITS-90, CVD, IEC-751, resistance
Reference measurement capability4 wire
Reference probe connection6-pin din with INFO-CON technology
Built-in RTD thermometer readout accuracyNI-120: ±0.015°C at 0°C
RTD resistance range0 Ω to 400 Ω
Resistance accuracy40 Ω to 25 Ω: ±0.002 Ω
RTD characterizationsPT-100 (385), (JIS), (3926), NI-120, resistance
RTD measurement capability2-wire, 3-wire, and 4-wire RTD with jumpers only
RTD connection4-terminal input
Built-in TC thermometer readout accuracy5Type J: ±0.70°C at 140°C
TC millivolt range–10 mV to 75 mV
Voltage accuracy0.025% of reading +0.01 mV
Internal cold junction compensation accuracy±0.35°C (ambient of 13°C to 33°C)
TC connectionMiniature connectors (ASTM E1684)
Built-in mA readout accuracy0.02% of reading + 0.002 mA
mA rangeCal 4-22 mA, Spec 4-24 mA
mA connection2 terminal input
Loop power function24 V DC loop power

Properties

MOQ1

Cross-References

Attached Documents