| Manufacturer part number (MPN) | 2452080 Brand: Fluke Category: Tools Voltage DC 30.000 V: 0.01% + 2 mV (upper display) Voltage DC 20.000 V: 0.01% + 2 mV (lower display) Voltage DC 90.00 mV: 0.01% + 10 μV Voltage DC -10.00 mV to 75.00 mV: 0.025% + 1 count (via TC connector) Current DC 24.000 mA: 0.01%+ 2 mA Resistance 0.0 to 4000.0 Ω: ± 0.015% Frequency 2.0 to 1000.0 CPM: 0.05% + 0.1 cpm Frequency 1.0 to 1000.0 Hz: 0.05% + 0.1 Hz Frequency 1.00 to 15.00 kHz: 0.05% + 0.01 kHz Frequency Sensitivity: 1 V peak-to-peak-minimum Pressure: Accuracy from 0.025% of range using any of (50) 750P pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure |
| Voltage DC 100.00 mV | 0.01% + 10 μV Current DC 24.000 mA (Source): 0.01% + 2 μA Current DC 24.000 mA (Simulate): 0.01% + 2 μA Frequency 1.0 to 10.0 kHz: 0.25% Frequency 10.0 to 15.0 kHz: 0.5% Frequency Waveform: 5 V p-p squarewave, -0.1 V offset Measure and source accuracy Cu10: 1.8°C Measure and source accuracy NI-120: 0.15°C Measure and source accuracy PT-100 (385): 0.45°C Measure and source accuracy PT-100 (392): 0.35°C Measure and source accuracy PT-100 (391-JIS): 0.35°C Measure and source accuracy PT-200 (385): 0.95°C Measure and source accuracy PT-500 (385): 0.55°C Measure and source accuracy PT-1000 (385): 0.35°C Measure and source accuracy RTD Resolution: 0.01°C (0.01°F) Measure and source accuracy Cu10.01 C: 0.1°C (0.1°F) Measure and source accuracy J: 0.4°C Measure and source accuracy K: 0.5°C Measure and source accuracy T: 0.4°C Measure and source accuracy E: 0.4°C Measure and source accuracy R: 1.4°C Measure and source accuracy S: 1.4°C Measure and source accuracy B: 1.4°C Measure and source accuracy L: 0.45°C Measure and source accuracy U: 0.45°C Measure and source accuracy N: 0.6°C Measure and source accuracy XK: 0.4°C Measure and source accuracy BP: 1.1°C Measure and source accuracy TC Resolution: J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F Ramp functions Source functions: Voltage, current, resistance, frequency, temperature Ramp functions Ramps: Slow ramp, Fast ramp, 25% step-ramp Loop power function Voltage: 24 V Loop power function Accuracy: 10% Loop power function Maximum current: 22 mA, short circuit protected Step functions Source functions: Voltage, current, resistance, frequency, temperature Step functions Steps: 25% of range, 100% of range Operating temperature: -10°C to +55°C Storage temperature: -20°C to +71°C Operating altitude: 3000 m Agency approvals: EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 Size: 200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in) Weight: 650 g (23 oz) Batteries: 4 AA alkaline batteries Battery replacement: Separate battery compartment, accessible without breaking calibration seal Side port connections: Pressure module connector, also used for remote real-time programming |