| Manufacturer part number (MPN) | 664133 Brand: Fluke Category: Tools Voltage DC 30.000 V: 0.02% + 2 counts (upper display) Voltage DC 100.00 mV: 0.02% + 2 counts Voltage DC -10.00 mV to 75.00 mV: 0.025% + 1 count (via TC connector) Current DC 24.000 mA: 0.02% + 2 counts Resistance 0.0 to 400.0 Ω: 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) Resistance 401 to 1500 Ω: 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) Resistance 1500 to 3200 Ω: 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) Frequency 2.0 to 1000.0 CPM: 0.05% + 1 count Frequency 1.0 to 1100.0 Hz: 0.05% + 1 count Frequency 1.00 to 10.00 kHz: 0.05% + 1 count Frequency Sensitivity: 1 V peak-to-peak-minimum Pressure: Accuracy from 0.025% of range using any of (50) 750P pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure |
| Voltage DC 10.000 V | 0.02% + 2 counts Current DC 24.000 mA (source): 0.02% + 2 counts Current DC 24.000 mA (simulate): 0.02% + 2 counts Resistance 15.0 to 400.0 Ω: 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) Frequency Waveform: 5 V p-p squarewave, -0.1 V offset Measure accuracy NI-120: 0.2°C Measure accuracy PT-100 (385): 0.33°C Measure accuracy PT-100 (393): 0.3°C Measure accuracy PT-100 (JIS): 0.3°C Measure accuracy PT-200 (385): 0.2°C Measure accuracy PT-500 (385): 0.3°C Measure accuracy PT-1000 (385): 0.2°C Measure accuracy Resolution: 0.1°C Measure accuracy J: 0.7°C Measure accuracy K: 0.8°C Measure accuracy T: 0.8°C Measure accuracy E: 0.7°C Measure accuracy R: 1.8°C Measure accuracy S: 1.5°C Measure accuracy B: 1.4°C Measure accuracy L: 0.7°C Measure accuracy U: 0.75°C Measure accuracy N: 0.9°C Measure accuracy XK: 0.6°C Measure accuracy BP: 1.2°C Source accuracy NI-120: 0.2°C Source accuracy PT-100 (385): 0.33°C Source accuracy PT-100 (393): 0.3°C Source accuracy PT-100 (JIS): 0.3°C Source accuracy PT-200 (385): 0.2°C Source accuracy PT-500 (385): 0.3°C Source accuracy PT-1000 (385): 0.2°C Source accuracy Resolution: 0.1°C Resolution: J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C XK: 0.6°C BP: 1.2°C Ramp functions Source functions: Voltage, current, resistance, frequency, temperature Ramp functions Ramps: Slow ramp, Fast ramp, 25% step-ramp Loop power function Voltage: 24 V Loop power function Accuracy: 10% Loop power function Maximum current: 22 mA, short circuit protected Step functions Source functions: Voltage, current, resistance, frequency, temperature Step functions Steps: 25% of range, 100% of range Operating temperature: -10°C to +55°C Storage temperature: -20°C to 71°C Operating altitude: 3000 m Agency approvals: EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 Size: 130 x 236 x 61 mm Weight: 0.65 kg Batteries: 4 AA alkaline batteries Battery replacement: Separate battery compartment, accessable without breaking calibration seal Side port connections: Pressure module connector, also used for remote real-time programming |