| Manufacturer part number (MPN) | 4692716 |
| Brand | Fluke |
| Category | Tools |
| 600.0 mV Resolution | 0.1 mV |
| 600.0 mV Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) |
| 600.0 mV Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3) |
| 6.000 V Resolution | 0.001 V |
| 6.000 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(1% + 3) |
| 6.000 V Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3) |
| 1000 V Resolution | 1 V |
| 1000 V Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts) | ±(2% + 3) |
| 1000 V Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts) | ±(2% + 3)¹ |
| 600.0 mV Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) |
| 6.000 V Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) |
| 1000 V Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts) | +(2% + 3), -(6% - 3) |
| Range Resolution | Accuracy ±(% of Rdg + Counts) |
| 6.000 V dc 0.001 V | ±(0.09% + 2) |
| 60.00 V dc 0.01 V | ±(0.09% + 2) |
| 600.0 V dc 0.1 V | ±(0.09% + 2) |
| 1000 V dc 1 V | ±(0.09% + 2) |
| Input impedance | 10 MΩ (nominal), <100 pF |
| Normal mode rejection ratio | >60 dB @ 50 Hz or 60 Hz |
| Common mode rejection ratio | >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance) |
| 600.0 mVdc 0.1 mV | ±(0.1% + 1) |
| Range | 400 mA |
| Resolution | 0.1 mA |
| Accuracy ±(% of Rdg+Counts) | ±(1.5% + 2)¹ |
| Burden voltage (Typical) | 2 mV/mA |
| Overload | 600 mA for 2 minutes maximum |
| Fuse protection for mA input | 0.44 mA, 1000 V, IR 10 kA |
| AC conversion | Inputs are ac-coupled and calibrated to the rms value of its wave input |
| 600.0 Ω 0.1 Ω | ±(0.9% + 2) |
| 6.000 kΩ 0.001 kΩ | ±(0.9% + 2) |
| 60.00 kΩ 0.01 kΩ | ±(0.9% + 2) |
| 600.0 kΩ 0.1 KΩ | ±(0.9% + 2) |
| 6.000 MΩ 0.001 MΩ | ±(0.9% + 2) |
| 50.0 MΩ ² 0.01 MΩ | ±(1.5% + 3) |
| Overload protection | 1000 V rms or dc |
| Open circuit test voltage | <8.0 V dc |
| Short circuit current | <1.1 mA |
| Diode test indication | Display voltage drop: 0.6 V at1.0 mA nominal test current: |
| Accuracy | ±(2% + 3) |
| Continuity indication | Continuous audible tone for testresistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω |
| Open circuit voltage | <8.0 V |
| Response time | >1 m sec |
| 99.99 Hz 0.01 Hz | ±(0.1% + 1) |
| 999.9 Hz 0.1 Hz | ±(0.1% + 1) |
| 9.999 kHz 0.001 kHz | ±(0.1% + 1) |
| 99.99 kHz 0.01 kHz | ±(0.1% + 1) |
| 600.0 mV ac V ac Sensitivity (RMS Its Wave)¹ 5 Hz to 20 kHz | 100.0 mV |
| V ac Sensitivity (RMS Its Wave)¹ 20 kHz to 100 kHz | 150.0 mV |
| 6.0 V V ac Sensitivity (RMS Its Wave)¹ 5 Hz to 20 kHz | 1.0 V |
| 6.0 V V ac Sensitivity (RMS Its Wave)¹ 20 kHz to 100 kHz | 1.5 V |
| 6.0 V DC Trigger Levels¹ to 20 kHz² | -400.0 mV and 2.5 V |
| 60.0 V V ac Sensitivity (RMS Its Wave)¹ 5 Hz to 20 kHz | 10.0 V |
| 60.0 V V ac Sensitivity (RMS Its Wave)¹ 20 kHz to 100 kHz | 36.0 V |
| 60.0 V DC Trigger Levels¹ to 20 kHz² | 1.2 V and 4.0 V |
| 600.0 V V ac Sensitivity (RMS Its Wave)¹ 5 Hz to 20 kHz | 100.0 V |
| 600.0 V DC Trigger Levels¹ to 20 kHz² | 12.0 V and 40.0 V |
| 1000.0 V V ac Sensitivity (RMS Its Wave)¹ 5 Hz to 20 kHz | 300.0 V |
| 1000.0 V DC Trigger Levels¹ to 20 kHz² | 12.0 V and 40.0 V |
| 1000 nF 1 nF | ±(1.2% + 2) |
| 10.00 μF 0.01 μF | ±(1.2% + 2) |
| 100.0 μF 0.1 μF | ±(1.2% ±90 counts) |
| 9999 μF 1 μF | ±(1.2% ±90 counts) |
| -40 ° C to 537 ° C 0.1 °C | ±(1% + 10 counts) |
| -40 ° F to 998 ° F 0.1 °F | ±(1% + 18 counts) |
| Measurement range | 0.01 MΩ to 2 GΩ |
| Test voltages | 50, 100, 250, 500, 1000 V |
| Test voltage accuracy | +20%, -0% |
| Short-circuit test current | 1 mA nominal |
| Auto discharge | Discharge time <0.5 second for C = 1 μF or less |
| Live circuit detection | Inhibit test if terminal voltage > 30 V prior to initialization of test |
| Maximum capacitive load | Operable with up to 1 μF load |
| 50 V (0% to +20%) Display range | 0.01 to 6.00 MΩ |
| 50 V (0% to +20%) Resolution | 0.01 MΩ |
| 50 V (0% to +20%) Test current | 1 mA @ 50 kΩ |
| 50 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts) | ±(3% + 5 counts) |
| 100 V (0% to +20%) Display range | 0.01 to 6.00 MΩ |
| 100 V (0% to +20%) Resolution | 0.01 MΩ |
| 100 V (0% to +20%) Test current | 1 mA @ 100 kΩ |
| 100 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts) | ±(3% + 5 counts) |
| 250 V (0% to +20%) Display range | 0.1 to 60.0 MΩ |
| 250 V (0% to +20%) Resolution | 0.1 MΩ |
| 250 V (0% to +20%) Test current | 1 mA @ 250 kΩ |
| 250 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts) | ±(1.5% + 5 counts) |
| 500 V (0% to +20%) Display range | 0.1 to 60.0 MΩ |
| 500 V (0% to +20%) Resolution | 0.1 MΩ |
| 500 V (0% to +20%) Test current | 1 mA @ 500 kΩ |
| 500 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts) | ±(1.5% + 5 counts) |
| 1000 V (0% to +20%) Display range | 0.1 to 60.0 MΩ |
| 1000 V (0% to +20%) Resolution | 0.1 MΩ |
| 1000 V (0% to +20%) Test current | 1 mA @ 1 MΩ |
| 1000 V (0% to +20%) Resistance accuracy ±(% of Rdg + Counts) | ±(1.5% + 5 counts) |
| Maximum voltage applied to any terminal and common | 1000 V |
| Storage temperature | -40 °C to 60 °C (-40 °F to 140 °F) |
| Operating temperature | -20 °C to 55 °C (-4 °F to 131 °F) |
| Temperature coefficient | 0.05 x (specified accuracy) per °C for temperatures 28 °C (82 °F) |
| Relative humidity | Noncondensing |
| Vibration | Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument |
| Radio frequency communication | 2.4 GHz ISM Band |
| Radio frequency certification | FCC: T68-FBLE |
| IC | 6627A-FBLE |
| International | IEC 61326-1:Portable Electromagnetic Environment; IEC 61326-2-2 CISPR 11: Group 1, Class A |
| Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) |
| USA (FCC) | 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. |
| Enclosure protection | IEC 60529: IP40 (non-operating) |
| IEC 61010-1 | Pollution Degree 2 |
| IEC 61010-2-033 | CAT IV 600 V / CAT III 1000 V |
| Batteries | Four AA batteries (NEDA 15A or IEC LR6) |
| Battery life | Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off. |
| Size (H x W x L) | 5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in) |
| Weight | 550 g (1.2 lb) |
| Altitude Operating | 2000 m |
| Altitude Storage | 12,000 m |
| Over-range capability | 110% of range except for capacitance which is 100% |
| Frequency overload protection | <107 V Hz |
| Nominal current range | 400 A |
| Continuous current range | 1 A - 400 A |
| Maximum non-destructive current | 1000 A |
| Lowest measurable current | 1 A |
| Basic accuracy | 2% + 0.06 A |
| Useable frequency | 5 Hz - 20 kHz |
| Output level(s) | 1 mA/A |
| Safety | CAT IV 600 V, CAT III, 1000 V |
| Maximum voltage | 1000 V AC rms from earth to ground, in compliance with EN61010 |
| Warranty | 1 year |
| Maximum conductor diameter | 32 mm (1.25 in) |
| Maximum conductor size (busbar) | 750 MCM |
| Output cable length | 1.5 m (59 in) |
| Shrouded banana plugs | Yes |
| Voltage range | 40-700 V |
| Frequency range | 15-400 Hz |
| Operating timer | Continuous |
| Dust/water resistance | IP 40 |
| Electrical safety | EN 61010, EN 61557-7 |
| Overvoltage category | CAT III, 600 V |
| Size | 124 x 61 x 27 mm |
| Power supply | From unit under test |