| Manufacturer part number (MPN) | 4489610 |
| Brand | Fluke |
| Category | Tools |
| Frequency specifications | Level specifications |
| Resolution 10 µHz | 0.001 dB |
| Frequency accuracy | ± 0.05 ppm ± 5 µHz |
| Attenuation accuracy (into 50 Ω) | Leveling Head O/P: ± 0.02 dB to 49 dB, ± 0.15 dB at 110 dB Relative to +10 dBm, 10 Hz to 128 MHz |
| VSWR | Leveling Head O/P (50 Ω): ≤ 100 MHz: ≤ 1.05, ≤ 2 GHz: ≤ 1.1, 2 GHz to 4 GHz: ≤ 1.0 + 0.05 x f (GHz) Microwave O/P: direct ≤ 2.0 (typical), HF option splitter device ≤ 1.22 |
| Harmonics and sub-harmonics | Harmonics ≤1 GHz: -60 dBc, >1 GHz: -55 dBc; sub-harmonics ≤ 4 GHz: none, >4GHz: -60 dBc |
| Spurious ≥ 3 kHz offset | ≤ -84 dBc at 500 MHz, ≤ -78 dBc at 1 GHz, ≤ -66 dBc at 4 GHz, ≤ -48 dBc at 27 GHz |
| Phase noise at 1 GHz | –144 dBc/Hz, typical, at 10 kHz to 100 kHz offset |
| Modulation | AM, FM, PM, internal and external to 4 GHz (Leveling Head and Microwave O/P); frequency pull and external leveling |
| Frequency sweep | 1 mHz to 4 GHz (Leveling Head); 1 mHz to 27 GHz (Microwave O/P). Linear or logarithmic. Stop-start or center-span |
| Frequency counter | Internal 300 MHz frequency counter |
| Power meter readout compatibility | Rhode & Schwarz NRP-Z series thermal sensor models 51, 52, 55.03, 55.04, 56, 57 and 58 |
| Temperature | Operating: 0 °C to 50 °C, 23 °C ± 5 °C for specified performance Storage: -20 °C to +70 °C |
| Standard interfaces | IEEE-488.2 (GPIB) |
| GPIB command emulation | 9640A, 9640A-LPN, 9640A-LPNX, HP3335, HP8662A, HP8663A, HP8340A, HP8360 B-Series; Agilent E8257 Series |
| Dimensions (HxWxD) | 146 mm x 442 mm x 551 mm (5.8 in x 17.4 in x 21.7 in) including handles. Industry-standard 19 in (483 mm) rack mounting when fitted with Y9600 rack mounting kit |
| Weight | 18 kg (40 lb) |