| Manufacturer part number (MPN) | 4922025 |
| Brand | Fluke |
| Category | Tools |
| Power requirements | 100 V ac to 240 V ac, 47 Hz to 63 Hz |
| Fuse | T 2 A, 250 V |
| Max power consumption | 100 W |
| Operating ambient temperature range | 15 °C to 35 °C |
| Storage temperature | -20 °C to 70 °C |
| Relative humidity | Operating: <80 % to 30 °C, <70 % to 40 °C |
| Vibration | MIL-T-28800E |
| Altitude (operation) | <3000 m |
| Warm up time | 15 minutes after power up or module installation, when items previously stored within Operating Ambient Temperature Range. |
| Ingress protection | IEC 60529:IP20 |
| Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
| Weight (Chassis only) | 13 kg (28.5 lbs) / 15 kg (35.25 lbs) |
| Dimensions | Height: 147 mm (5.78 in) |
| Rack mount dimensions | 3U-19 inch rack |
| USA (FCC) | 47 CFR 15 subpart B, this product is considered an exempt device per clause 15.103. |
| Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes. |
| Primary Remote Interfaces | IEEE, Ethernet, RS-232, USB |
| System Connection | Supports interconnection of 2 or 3 systems |
| Switch Test Connection | Standard BNC Jack: Nominal 24 V dc isolated drive Maximum 30 V dc w.r.t.chassis ground |
| Aux Drivers | 4 external Solenoid Drivers 24 V dc Drive (Maximum drive 6 W continuous per channel |
| Control Precision (Dynamic Mode) | 0.002% Range Span or 0.01 kPa (whichever is greater) |
| Control Turndown | 100:1 (typical) To meet the control specifications, supply pressure should not be greater than 100 times the range of the measurement module. Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range |
| Low Control Point | 1 kPa (0.15 psi) absolute (8270A only) 7 kPa (1.0 psi) gauge (8370A, 8270A without vacuum pump) |
| Range Range > 44 MPa | Range |
| PM200/PM230 30 seconds | 45 seconds |
| PM600/PM630; 45 seconds | 60 seconds |