752A Reference Divider

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
590539
752A Reference Divider

752A reference divider compensates for long-term changes in value of the divider resistors by switching positions in various Wheatstone bridge configuration

Technical Specifications

Manufacturer part number (MPN)590539
BrandFluke
CategoryTools
Ratio Ranges10:1 and 100:1
Temperature Coefficient£ ± 1 ppm per °C over range of 18°C to 28°C (typically 0.1 ppm per °C from 15°C to 30°C
Input Resistance10:1 Ratio:380 kΩ ± 1% 100:1 Ratio:Divider is 4 MΩ; Driven Guard is 4 MΩ; total is 2 MΩ ± 1%
Maximum Input Voltage200V for the 10:1 ratio 1100V for the 100:1 ratio
Power Coefficient£ 0.05 ppm of output with 100V applied for 10:1 ratio and £ 0.3 ppm of output with 1000V applied for 100:1 ratio (included in the ratio accuracy specifications)
Temperature0°C to 50°C, operating; -40°C to 75°C non-operating
Relative Humidity£ 75% to 40°C, £ 45% to 50°C, non-condensing, operating; £ 100% 10-50°C, non-operating
Altitude£ 3050m (10,000 ft) operating;£ 12,220m (40,000 ft) non-operating
VibrationPer MIL-T-28800C, Type III, Class 5, Style E
SafetyIEC 348, 2nd edition, 1978; ANSI-C39.5, 1980, CSA 556B, and UL 1244
Calibration DocumentationStatement of calibration practices included; performance check optional
Size19.1 cm H x 22.1 cm W x 60.3 cm L (7.53 in H x 8.69 in W x 23.75 in L)
Weight8.4 kg (18.5 lbs.)

Properties