cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

5901 Triple Point of Water Cells

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
2507466
5901 Triple Point of Water Cells

5901 triple point water cells from Fluke Calibration are an easy-to-use and inexpensive standard with uncertainty better than ± 0.0001 °C.

Technical Specifications

Manufacturer part number (MPN)2507466
BrandFluke
CategoryTools
Specifications 5901A-G5901A-Q
Expanded Uncertainty (k=2) < 0.0001 °C< 0.0002 °C
Repro- ducibility 0.00002 °C0.00005 °C
Dimensions 50 mm OD 12 mm ID 450 mm long60 mm OD 13.6 mm ID 420 mm long
Immersion Depth (water surface to well bottom) 265 mm118 mm
Material Borosilicate GlassFused Silica (Quartz)
Water SourceOcean
dDVSMOW ± 10 ‰ (± 1 %)± 20 ‰
d18OVSMOW ± 1.5 ‰ (± 0.15 %)± 3 ‰
Effect of deviation from VSMOW ± 7 µK± 14 µK

Properties

MOQ1

Cross-References

Attached Documents