SPRT from Fluke Calibration are highly stable, have low drift rates, and can be used over a wide range of temperature. Request a product quote today.
| Manufacturer part number (MPN) | 1656319 |
| Brand | Fluke |
| Category | Tools |
| Temperature Range –200 °C to 670 °C | –200 °C to 480 °C |
| Nominal RTPW 25.5 Ω | 0.25 Ω |
| Current 1 mA | 14.14 mA |
| W(302.9146 K) ≥ 1.11807 and W(234.3156 K) ≤ 0.844235 | W(302.9146 K) ≥ 1.11807 and W(1234.93 K) ≥ 4.2844 |
| Sensitivity 0.1 Ω/°C | 0.001 Ω/°C |
| < 0.002 °C/100 hours at 661 °C (typically < 0.001 °C) | < 0.001 °C/100 hours at 480 °C (0.0005 °C typical) |
| Diameter of Sensor Pt Wire 0.07 mm (0.003 in) | 0.4 mm (0.016 in) |
| MOQ | 1 |