The Fluke Calibration 5606 immersion PRT probe offers excellent accuracy and stability for temperature measurement from -200°C to 160°C.
| Manufacturer part number (MPN) | 3622907 Brand: Fluke Category: Tools Temperature range: -200°C to 160°C Nominal resistance at 0.01°C: 100 Ω ±0.1 Ω Temperature coefficient: 0.00385 Ω/Ω/°C Accuracy1: ±0.04°C at 0.01°C ±0.06°C at 160°C Short-term repeatability2: ±0.03°C at 0.01°C ±0.04°C at 160°C Drift3: ±0.03°C at 0.01°C ±0.04°C at 160°C Hysteresis: ±0.015°C Sheath length: 50 mm ±5 mm (2 in ±0.2 in) Sheath diameter: 3.1 mm ±0.1 mm (1/8 in ±0.004 in) Sheath material: 316 SST Transition junction temperature range4: -200°C to 160°C Transition junction dimensions: No transition Sensor length: 30 mm ±3mm (1.2 in ±0.1 in) Sensor location: 3 mm ±1 mm from tip (1.2 in ±0.04 in) minimum insulation resistance: 20 MΩ at 23°C minimum immersion length: Full immersion Maximum immersion depth in liquid medium: Full immersion Maximum immersion depth in dry Medium: Full immersion Response time5: 12 seconds typical Self heating (in 0°C bath): ±0.003°C Lead-wire cable type: Enameled copper wire Lead-wire length: 2.4 meters (8 ft) Lead-wire temperature range: 160°C Temperature -197°C: 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.4°C -38°C: 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.2°C 0°C: 1924 Calibration Uncertainty6 0.03°C Tolerance w/out Calibration7 0.1°C 100°C: 1924 Calibration Uncertainty6 0.045°C Tolerance w/out Calibration7 0.3°C 157°C: 1924 Calibration Uncertainty6 0.05°C Tolerance w/out Calibration7 0.4°C |
| MOQ | 1 |