cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

52120A Transconductance Amplifier

Fluke · Safety & Detection · Tools · USA · fluke.com
Part Number
4044885
52120A Transconductance Amplifier

Fluke Calibration 52120A Transconductance Amplifier supplies dc currents up to 100 amps & ac currents up to 120 amps, with accuracies up to 140 ppm.

Technical Specifications

Manufacturer part number (MPN)4044885
BrandFluke
CategoryTools
Output ranges2 A, 20 A, 120 A (100 A dc)
Input current range200 mA, 200 mA, 120 mA (100 mA dc)
Current gain10, 100, 1000
Input voltage range2 V, 2 V, 1.2 V (1.0 V dc)
Transconductance1,10,100 Siemens
FrequencyTo 10 kHz
Maximum output compliance voltage4.5 Vrms (6.4 Vpeak)
Output isolation, current terminal to earth600 Vrms, 850 Vpeak, dc to 850 Hz
AC accuracy, closed loop with 6105ATo 140 ppm
DC accuracy, standaloneTo 160 ppm
AC accuracy, standaloneTo 350 ppm
Phase angle accuracy0.006°
Load dependent phase angle shift<0.001° @ 60 Hz
Input line voltage range100 V to 240 V with up to ±10 % fluctuations
Transient overvoltageImpulse withstand (overvoltage); Category II of IEC 60364-4-443
Maximum consumption< 1500 VA
Calibration DocumentationReport of calibration with data included; 17025 accredited report optional
Dimensions with feet (H x W x L)192 mm x 432 mm x 645 mm (7.6 inches x 17 inches x 25.5 inches)
Dimensions without feet (H x W x L)178 mm x 432 mm x 645 mm (7 inches x 17 inches x 25.5 inches)
Weight25 kg (54 lb.)
Design standards and complianceDesigned to EN 61010-1: 2010, CAN/CSA 22.2 No 61010.1-04, ANSI/UL 61010-1:2004, EN 61326-1:2006 CE marked, CSA listed
Operating temperature5 °C to 35 °C
Calibration temperature range16 °C to 30 °C
Warm up time1 hour
Safe operating max. relative humidity (non-condensing)<80 % 5 °C to 31 °C ramping linearly down to 50 % at 35 °C
Operating altitude0 m to 2,500 m
Shock and vibrationMIL-PRF-28800F class 3
Specification confidence level99%

Properties

MOQ1

Cross-References

Attached Documents