Danfoss MBT 9110 temperature transmitter, code 084Z8119 (type designation MBT 9110-000B-D125). Two-wire loop-powered transmitter that converts a Pt 100 resistance input into a linear 4-20 mA output over a factory-set measuring range of 0 to +250 C (+32 to +482 F). Pt 100 platinum input (100 ohm at 0 C) per EN 60751. Supply voltage 8-35 V DC; 2-wire connection, 2 terminals. IP54 enclosure; standard box (head-mount) housing. Supplied in an 8-piece multipack. CE marked.
| Type | MBT 9110 (temperature transmitter) |
| Type designation | MBT 9110-000B-D125 |
| Code number | 084Z8119 |
| Sensor input | Pt 100 platinum resistance (100 ohm at 0 C) per EN 60751 |
| Measuring range | 0 to +250 C (+32 to +482 F) |
| Output signal | 4-20 mA (2-wire, loop powered) |
| Supply voltage | 8-35 V DC |
| Electrical connection | 2-wire output, 2 terminals |
| Enclosure rating | IP54 |
| Transmitter type | Standard in box (head-mount) |
| Packing format | Multi pack (8 pc) |
| Approvals | CE |
| EAN | 5702423098712 |
| Net weight | 0.394 Kilogram |
| Product group | Sensors |
| Ambient temperature max. [°C] | 85 °C |
| Ambient temperature max. [°F] | 185 °F |
| Approval | CE |
| CE mark | Yes |
| China RoHS compliance | Out of scope |
| EEE category | 5 small equipment (any external dimension < 50 cm) |
| Electrical connection size | 2 terminals |
| Enclosure rating IP | IP54 |
| EU RoHS compliance | Yes with exemptions |
| EU RoHS exemption clause | 6(c) |
| In scope of WEEE | Yes |
| Insertion type | Fixed |
| Measuring range [°C] [max] | 250 °C |
| Measuring range [°C] [min] | 0 °C |
| Measuring range [°F] [max] | 482 °F |
| Measuring range [°F] [min] | 32 °F |
| Number of terminals [pcs] | 2 pc |
| Output signal [mA] [max] | 20 mA |
| Output signal [mA] [min] | 4 mA |
| REACH Candidate List substances | Lead (CAS no. 7439-92-1) |
| Sensor elem. resistance [Ohm] | 100 Ohm |
| Sensor element [pc] | 1 pc |
| Sensor element type | Pt |
| Supply voltage [V] DC [max] | 35 V |
| Supply voltage [V] DC [min] | 8 V |
| Temperature range [°C] [max] | 850 °C |
| Temperature range [°C] [min] | -200 °C |
| Temperature range [°F] [max] | 1562 °F |
| Temperature range [°F] [min] | -328 °F |
| Transmitter setting range [°C] [max] | 250 °C |
| Transmitter setting range [°C] [min] | 0 °C |
| Transmitter setting range [°F] [max] | 482 °F |
| Transmitter setting range [°F] [min] | 32 °F |
| Wires [pcs] | 3 pc |
| Temperature range | 0 to +250 C |
| IP rating | IP54 |
| MOQ | 1 |
MBT 9110 loop-powered transmitter with a factory-set 0 to +250 C range. Pair with a Pt 100 sensor (e.g. an MBT 5250) and confirm the range suits the process and the 4-20 mA loop before ordering; range is not field-adjustable.