cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

Danfoss MBT 5560 temperature transmitter, Pt 1000, 4-20 mA (0 to +100 C), 250 mm insertion, G 1/4 process connection, PG9 angular (code 084Z4086)

Danfoss · Sensors & Instrumentation · Temperature Sensors · Denmark · danfoss.com
Part Number
084Z4086  ·  HS 9025.19
Danfoss MBT 5560 temperature transmitter, Pt 1000, 4-20 mA (0 to +100 C), 250 mm insertion, G 1/4 process connection, PG9 angular (code 084Z4086)

Danfoss MBT 5560 insertion temperature transmitter, code 084Z4086 (type designation MBT 5560-0001-0250-10-110). Compact temperature sensor with an integrated 2-wire, loop-powered transmitter that converts the resistance of a built-in Pt 1000 platinum element into a standardized 4-20 mA signal, so a defined temperature is transmitted directly without a separate head or rail transmitter. For engine cooling water, lubricating and hydraulic oil, compressor, refrigeration and HVAC/R circuits, and marine and industrial plant. Insertion length 250 mm; G 1/4 male (ISO 228-1-A) process connection; AISI 316 Ti stainless-steel protection tube, 8 mm dia. Element measuring range -50 to +200 C (-58 to +392 F). Output 4-20 mA (2-wire) over a 0 to +100 C span; supply 10-32 V DC. Electrical connection: Angular connector, Pg 9 (EN 175301-803-A), male and female. IP65 enclosure per IEC 60529. The built-in transmitter is factory-scaled to a 0 to +100 C output span. Supplied in a 9-piece multipack. CE marked.

Technical Specifications

TypeMBT 5560 temperature sensor with built-in transmitter
Type designationMBT 5560-0001-0250-10-110
Code number084Z4086
Sensor elementPt 1000 platinum resistance (1000 ohm at 0 C), EN 60751
Transmitter accuracy< ±1% FS
Measuring range (element)-50 to +200 C (-58 to +392 F)
Output signal4-20 mA (2-wire, loop-powered), transmitter setting range 0 to +100 C
Supply voltage10-32 V DC
Insertion length250 mm
Process connectionG 1/4 male (ISO 228-1-A)
Protection tubeAISI 316 Ti stainless steel, 8 mm dia.
Electrical connectionAngular connector, Pg 9 (EN 175301-803-A), male and female
Enclosure ratingIP65 per IEC 60529
Packing quantity9 pc (multipack)
Net weight0.162 kg
ApprovalsCE
EAN5702423239542
Product groupSensors
Ambient temperature max. [°C]85 °C
Ambient temperature max. [°F]185 °F
ApprovalCE
CE markYes
China RoHS complianceOut of scope
Connection materialAISI 316 Ti
Connection standardISO 228
Connection typeG
EEE category5 small equipment (any external dimension < 50 cm)
Electrical conn
Male/FemaleMale and Female
Electrical connection sizePg 9
Electrical connection standardEN 175301-803-A
Enclosure rating IPIP65
EU RoHS complianceYes with exemptions
EU RoHS exemption clause6(c)
Extension length [in]1.3 in
Extension length [mm]33 mm
In scope of WEEENo
Insertion length [in]9.8 in
Insertion length [mm]250 mm
Insertion typeFixed
Output signal [mA] [max]20 mA
Output signal [mA] [min]4 mA
Process connection Male/FemaleMale
Process connection materialStainless steel
Process connection size1/4
Process connection standardISO 228-1-A
Process connection typeG
Protection tube diameter [in]0.315 in
Protection tube diameter [mm]8 mm
Protection tube materialAISI 316 TI
REACH Candidate List substancesLead (CAS no. 7439-92-1)
Sensor element tolerance< ± 1% FS
Sensor InsertFixed
Supply voltage [V] DC [max]32 V
Supply voltage [V] DC [min]10 V
Temperature range [°C] [max]200 °C
Temperature range [°C] [min]-50 °C
Temperature range [°F] [max]392 °F
Temperature range [°F] [min]-58 °F
Transmitter setting range [°C] [max]100 °C
Transmitter setting range [°C] [min]0 °C
Transmitter setting range [°F] [max]212 °F
Transmitter setting range [°F] [min]32 °F
Transmitter typeIntegrated

Properties

MaterialAISI 316 Ti stainless steel
Temperature range-50 to +200 C (-58 to +392 F)
IP ratingIP65
Standard size250 mm insertion
Thread standardISO 228-1-A
MOQ1

Compatibility

MBT 5560 sensor with factory built-in transmitter — the 4-20 mA (2-wire) 0 to +100 C span is preset by the type designation and is not field-rangeable on the standard version. Confirm the required insertion length (250 mm), process thread (G 1/4), electrical connector and supply match the pocket/thermowell and controller input before ordering. A matching thermowell can be ordered separately.

Standards

Approvals

Attached Documents