cBallast
Industrial Parts Catalog
Datasheet generated 2026-10-07
sales@cballast.com
cballast.com

Bürkert Mass Flow Controller (MFC)/mass Flowmeter (MFM) for Gases Type 8744

Bürkert · Sensors & Instrumentation · Flow Meters · Germany · burkert.com
Part Number
Type 8744  ·  HS 9026.10
Bürkert Mass Flow Controller (MFC)/mass Flowmeter (MFM) for Gases Type 8744

The Type 8744 mass flow controller (MFC)/mass flow meter (MFM) is suitable for controlling or measuring aggressive gases, gas mixtures and in the area of application for alternating gases. A thermal capillary sensor, with non-wetted sensor elements, is used for this. This also enables easy conversion to the process gas in question. The wetted parts are made of high-quality stainless steel and FFKM as seal material for high chemical resistance. The MFC variant achieves high control accuracy thanks to the use of Bürkert proportional valve technology and the application-appropriate layout of the valve orifice. The MFC can be implemented with low pressure loss as required, due to the diversity of the available valve orifices. These MFCs and MFMs either communicate the set-point and actual values of the flow value via an analogue interface, or other values, as well as the set-point and actual values, that can be selected via software via an Industrial Ethernet interface. Type 8744 represents the IP65 variant of the Type 8743 and contains the CAN-based büS variant for integration into CAN or büS networks. A wide range of devices can easily and efficiently be integrated into the control level in this way via an Ethernet gateway. The communication data is configured via the Bürkert Communicator software.

Technical Specifications

SealFKM, EPDM and FFKM, PCTFE seat seal for nominal diameters DN 0.05 and DN 0.1
HousingAluminium die casting (painted)
Base blockStainless steel 1.4404/316L
Wetted partsStainless steel 1.4401/316, 1.4404/316L, 1.4435/316L, PCTFE and seal material
Total weightVariant for nominal flow rates < 20 l/min: Variant for nominal flow rates > 20 l/min:
LED displayRGB LED according to NAMUR NE107
Nominal flow range (QN)10 mlN/min…100 lN/min (N2)
Operating pressureMFM: max. 10 bar
Repeatability± 0.1 % FS
Turndown ratio1:20 at QN < 25 ml/min, 1:50 at QN > 25 ml/min (higher on request)
Temperature coefficient± (0.05 % FS + 0.05 % MV)/K (deviation from gas temperature during calibration and adjust-
Pressure coefficient± 0.1 % MV/bar (deviation from operating pressure during calibration and adjustment, referring
Settling time (MFC)/Response time (MFM)< 2 s
Actuator (proportional valve)Normally closed
Nominal diameter0.05…4 mm

Properties

MOQ1

Cross-References

Attached Documents